FN280
Compact filtered power entry module
Power Entry Module with EMC- Filter FN280
Compact filtered power entry module
- Complies with IEC/EN 60601-1
- Rated currents up to 10 A
- Single or dual-fuse holder
- Fuses Ã5 x 20 mm
- 2-pole rocker switch
- General purpose application
- Optional earth line choke (E type)
- Optional medical versions (B type)
Features and Benefits
High conducted attenuation performance, based on chokes with high saturation resistance and excellent thermal behavior
Rear/front or snap-in mounting
Single or dual-fuse holder
2-pole rocker switch
Custom-specific versions are available on request
Typical Applications
Portable electrical and electronic equipment
Consumer goods
EDP and office equipment
Single-phase power supplies, switch-mode power supplies
Test and measurement equipment
Medical equipment
Technical Specifications
Maximum continuous operating voltage
250 VOLT, 50/60 Hz
Operating frequency
DC to 400 Hz
High potential test voltage
P –> PE 2000 VAC for 2 sec (standard types)
P –> PE 2500 VAC for 2 sec (B types)
P –> N 760 VAC for 2 sec
Protection category
IP 40 according to IEC 60529
Temperature range (operation and storage)
-25°C to +85°C (25/85/21)
Design corresponding to
UL 1283, CSA 22.2 No. 8 1986, IEC/EN 60939
Flammability corresponding to
UL 94 V-2 or better
Fuse holder
1 or 2 fuses (Ø5 x 20 mm) (certified to IEC 60127-6), power acceptance 1.6W @ Ta 23°C per pole for 2 fuses. 2.5W, if only 1 pole is used.
Marking
) Marking max. 250 V ~ (text is print. below fuse symbol)
Mechanical life
50,000 cycles
Rocker switch description
 
Function
2-pole, dark not illuminated
Marking I – 0
Electrical specifications
Inrush current 82 A
6,000 on-off operations according to UL 1054, TV5
10,000 on-off operations according to ENEC
Switch ratings
 
USA (UL) and Canada (C-UL)
10 A, 125 VAC
10 A, 250 VAC
1/3 HP
Europe (ENEC)
10 A (4 A), 250 VAC*
Rated currents
1 to 10 A @ 40°C max